Collapsibility Testing Of Aluminium Tube

This test is performed with the help of a collapsibility tester instrument.

  • Place the empty aluminium tube on the groove on the top of the auxiliary base plate,
  • on the base of the apparatus with the open end touching the stopper,
  • when releasing the shutter and allowing it to fall upon the empty aluminium tube,
  • Read the reading on the measuring scale the position of the top of the shutter.
  • Acceptable Quality Levels: When the empty aluminium tube is testing this manner, the minimum and maximum limit for collapsibility reading we can use concerning the Tube diameter, is below the parameter table.

Diameter of Aluminium tube

Collapsibility Reading, MM

MIN

MAX

13.5

7

10

16

4

6

19

5

8

22

6

10

25

11

15

30

16

20

35

21

25

40

24

28

45

30

35